Description
Generator
- 60kHz high frequency
- 15kW power
- Up to 120kVp
- Up to 75mA for radiographic film exposure
- Continuous high level fluoro (HLF) up to 20mA
- Digital spot up to 75mA
- Digital cine pulse
- 15 or 30 pulses per second, 60 Hz
- 12 or 25 pulses per second, 50 Hz
- Up to 150mA
- 10ms pulse width
- Full powerfrom standard wall outlet
- Patented battery buffered design
X-ray Tube
- Rotating anode X-ray tube
- 0.3 and 0.6 nominal focal spots
- Anode heat capacity: 300,000 HU
- Anode cooling rate: 85,000 HU/min.
- Housing heat capacity: 1,600,000 HU
- Housing cooling rate:
- 9” I.I. Super C- 22,500 HU/min
- 12” I.I. – 22,500 HU/min
Digital Image Rotation
- Digitally adjusts image display
- Automatic image update
- Image rotation
- Image reversal (side-to-side)
- Image invert (top-to-bottom)
- Image positioning without additional exposure
Preview Collimator
- On-screen collimator position indication
- PreView iris collimator
- PreView Tungsten rotatable double leaf collimator
- Adjusts collimators without X-ray exposure
Fluoro Mode
- kVp range: 40 -120
- mA range:
- 0.2 – 10 normal mode
- 1.0 – 20 HLF (high level fluoro)
- Auto and manual fluoro modes
- AutoTrak ABS varies mA, kVp, camera gain
Pulsed Fluoro Mode
- kVp range: 40 – 120
- mA range: 0.2 – 10
- Pulse rate: 1, 2, 4, 8
- Pulse width: 25 or 50ms
- AutoTrak ABS, mA, kVp, camera gain
- Reduces X-ray dose to patient and operator
High Level Pulsed Fluoro
- kVp range: 40 – 120
- mA range: 1 – 40
- Pulse rate: 1, 2, 4, 8
- Pulse width: 25 or 50ms
- AutoTrak ABS, mA, kVp, and camera gain
Digital Spot Mode
- kVp range: 40 – 120
- mA range: Up to 75
- Automatic exposure termination and automatic image save
Radiographic Mode
- mA range: up to 75
- mAs range: up to 300
- Computer controlled exposure time
- Optional film cassette holder
- 10” x 12” (24cm x 30cm) for 9” I.I.
- 14” x 14” (35cm x 35cm) for 12” I.I.
Video Imaging System
9″ Image Intensifier
- Tri-mode 9”/6”/4.5” (23cm/15cm/11cm) image intensifier
- Minimum central resolution (at monitor):
- 9” (23cm): 2.2 lp/mm
- 6” (15cm): 3.0 lp/mm
- 4.5” (11cm): 3.5 lp/mm
- DQE: 65% (typical)
12″ Image Intensifier
- Tri-mode 12”/9”/6” (31cm/23cm/15cm) image intensifier
- Minimum central resolution (at monitor):
- 12” (23cm): 1.6 lp/mm
- 9” (15cm): 2.2 lp/mm
- 6” (11cm): 2.6 lp/mm
- DQE: 65% (typical)
Precision Imaging with Dynamic Range Management (DRM)
- Preset Imaging Profile: Orthopedic
- Enhances features of interest while attenuating background noise
AutoTrak Automatic Brightness Stabilization (ABS)
- Automatically seeks the subject anatomy anywhere within the imaging field and selects the optimum imaging technique
- Automatically adjusts to anatomical size and location
- Provides uniform image quality throughout the entire image
- Simplifies operation
Image Quality Features
- Smart Window
- Dynamically senses the collimator position and automatically adjusts brightness and contrast to produce high image quality
- Smart Metal
- Allows user to adjust automatic brightness and contrast sensitivity levels to metal
- Provides optimum image quality even when metal is introduced to the field
- Tungsten Collimator
- Dense collimator limits X-ray exposure area
- Reduces scatter radiation
- mproves image detail
Video Camera
- High resolution 1k x 1k CCD camera
- Full frame capture
- Motorized rotation
- On-screen orientation indicator (real-time feedback without fluoro)
- Left-right image reversal
- Top-bottom image invert
Video Display
- Dual 19” (48cm) display anti-glare, LCD flat panel monitors mounted on an articulating arm
- 22” horizontal travel
- 7° up/10° down
- Monitors viewable from all four sides of workstation
- Horizontal and vertical viewing angle 170°
- 1200 CD/M2 maximum brightness
- Touch screen system control
- 1280 x 1024 high resolution monitors
- Integrated color monitor for display of VGA, DVI, DVI-D, S-VHS, and SDISD formats
Brand
GE